Development of a New Pulsatile Ventricular Assist Device

https://hiroshima.repo.nii.ac.jp/records/2013573
https://hiroshima.repo.nii.ac.jp/records/2013573
06f848a9-3ad1-4289-8e09-067feb00f6f1
名前 / ファイル ライセンス アクション
HiroshimaJMedSci_55_29.pdf HiroshimaJMedSci_55_29.pdf (1.6 MB)
Item type デフォルトアイテムタイプ_(フル)(1)
公開日 2023-03-18
タイトル
タイトル Development of a New Pulsatile Ventricular Assist Device
言語 en
作成者 Kurosaki, Tatsuya

× Kurosaki, Tatsuya

en Kurosaki, Tatsuya

Search repository
Sakai, Hiroshi

× Sakai, Hiroshi

en Sakai, Hiroshi

Search repository
Ninomiya, Shinji

× Ninomiya, Shinji

en Ninomiya, Shinji

Search repository
Fukunaga, Shintaro

× Fukunaga, Shintaro

en Fukunaga, Shintaro

Search repository
Sueda, Taijiro

× Sueda, Taijiro

en Sueda, Taijiro

Search repository
アクセス権
アクセス権 open access
アクセス権URI http://purl.org/coar/access_right/c_abf2
権利情報
権利情報 (c) Hiroshima University Medical Press.
主題
主題Scheme Other
主題 Ventricular assist device
主題
主題Scheme Other
主題 Low cost
主題
主題Scheme Other
主題 Flow visualization
主題
主題Scheme NDC
主題 490
内容記述
内容記述 We developed a small, lightweight, low-cost implantable ventricular assist device (VAD) for use in smaller Japanese subjects. The major advantage of this pump is the simplicity of its fabrication. Most parts of the pump were shaped from a transparent acrylic block by a turning process, and the diaphragm was made from a silicon sheet. Since this method of construction did not require any complex processes, we could manufacture many pumps of various shapes. We determined the most efficient shape for the Ebacor VAD using the flow visualization technique. The pump showed an output above 6 liters/min under a driving pressure of 300/-100 mmHg. The pump performance of current VADs is superior to that of the Ebacor VAD, because these pumps are larger. Since the Ebacor VAD is small in size, it can be driven by the driving system of a normal IABP control unit, which many hospitals already have in place. During a 30day continuous driving performance test of this pump, no problems like performance decrements or water and air leakage were observed.
言語 en
出版者
出版者 Hiroshima University Medical Press
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ departmental bulletin paper
出版タイプ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
収録物識別子
収録物識別子タイプ ISSN
収録物識別子 0018-2052
収録物識別子
収録物識別子タイプ NCID
収録物識別子 AA00664312
開始ページ
開始ページ 29
書誌情報 Hiroshima Journal of Medical Sciences
Hiroshima Journal of Medical Sciences

巻 55, 号 1, p. 29-34, 発行日 2006-03
旧ID 34923
戻る
0
views
See details
Views

Versions

Ver.1 2025-02-23 06:37:15.202364
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3