Assessment of a New Silicone Tri-leaflet Valve Seamlessly Assembled with Blood Chamber for a Low-cost Ventricular Assist Device

https://hiroshima.repo.nii.ac.jp/records/2013429
https://hiroshima.repo.nii.ac.jp/records/2013429
30c8e3ac-6bde-4b4d-a9a8-37922b9c4ac2
名前 / ファイル ライセンス アクション
HiroshimaJMedSci_47_47.pdf HiroshimaJMedSci_47_47.pdf (1.4 MB)
アイテムタイプ デフォルトアイテムタイプ_(フル)(1)
公開日 2023-03-18
タイトル
タイトル Assessment of a New Silicone Tri-leaflet Valve Seamlessly Assembled with Blood Chamber for a Low-cost Ventricular Assist Device
言語 en
作成者 Hirai, Shinji

× Hirai, Shinji

en Hirai, Shinji

Search repository
Fukunaga, Shintaro

× Fukunaga, Shintaro

en Fukunaga, Shintaro

Search repository
Sueshiro, Masafumi

× Sueshiro, Masafumi

en Sueshiro, Masafumi

Search repository
Watari, Masanobu

× Watari, Masanobu

en Watari, Masanobu

Search repository
Sueda, Taijiro

× Sueda, Taijiro

en Sueda, Taijiro

Search repository
Matsuura, Yuichiro

× Matsuura, Yuichiro

en Matsuura, Yuichiro

Search repository
アクセス権
アクセス権 open access
アクセス権URI http://purl.org/coar/access_right/c_abf2
主題
主題Scheme Other
主題 Silicone tri-leaflet valve
主題
主題Scheme Other
主題 Silicone blood chamber
主題
主題Scheme Other
主題 Ventricular assist device
主題
主題Scheme NDC
主題 490
内容記述
内容記述 We have developed a practical, low-cost ventricular assist device (VAD) comprising a newly designed blood chamber with a silicone lenticular sac and two silicone tri-leaflet valves (STV), made en bloc. This new VAD is seamless, can be made cost-effectively and assembled with the blood chamber and valve as one body. This novel design should reduce the incident of thrombus formation because of the absence of a junction at the connecting ring and because of the use of flexible silicone materials which have both antithrombogenecity and biocompatibility. In in vitro hemodynamics testing, a batch of 3 consecutively manufactured V ADs with STV s underwent hydrodynamic functional testing. These showed less regurgitation, a lower value of water hammer phenomenon, and a slightly greater pressure gradient across the valves than a mechanical valve (MV) [Björk-Shiley monostrut valve]. The flow and pulsatile efficiency were adequate and similar to that of a VAD with MVs. In in vitro durability and hemolysis tests, a V AD with STV functioned well for 54 days and showed similar hemolytic profiles to a V AD with MVs. In an in vivo acute experiment using an adult sheep, our device was promblem-free providing sufficient output as a left ventricular assist device (LVAD). Although it will be necessary to decrease the pressure gradient across this STV in the future, our device showed efficient performance as a practical and cost-effective V AD for short term use.
言語 en
内容記述
内容記述タイプ Other
内容記述 This work was supported in part by the Tsuchiya Foundation and by a Research Grant for Cardiovascular Diseases (7A-1) from the Ministry of Health and Welfare, Japan
出版者
出版者 Hiroshima University Medical Press
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ departmental bulletin paper
出版タイプ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
収録物識別子
収録物識別子タイプ ISSN
収録物識別子 0018-2052
収録物識別子
収録物識別子タイプ NCID
収録物識別子 AA00664312
開始ページ
開始ページ 47
書誌情報 Hiroshima Journal of Medical Sciences
Hiroshima Journal of Medical Sciences

巻 47, 号 2, p. 47-55, 発行日 1998-06
旧ID 37792
戻る
0
views
See details
Views

Versions

Ver.1 2025-02-23 06:33:05.242512
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3