The Protective Effect of Low Power He-Ne Laser Against Erythrocytic Damage Caused by Artificial Heart-Lung Machines

https://hiroshima.repo.nii.ac.jp/records/2013385
https://hiroshima.repo.nii.ac.jp/records/2013385
6d5d7207-1be7-408f-8331-b2069ce221d1
名前 / ファイル ライセンス アクション
HiroshimaJMedSci_45_15.pdf HiroshimaJMedSci_45_15.pdf (929.7 KB)
Item type デフォルトアイテムタイプ_(フル)(1)
公開日 2023-03-18
タイトル
タイトル The Protective Effect of Low Power He-Ne Laser Against Erythrocytic Damage Caused by Artificial Heart-Lung Machines
言語 en
作成者 Takashi, Itoh

× Takashi, Itoh

en Takashi, Itoh

Search repository
Murakami, Hironobu

× Murakami, Hironobu

en Murakami, Hironobu

Search repository
Orihashi, Kazumasa

× Orihashi, Kazumasa

en Orihashi, Kazumasa

Search repository
Sueda, Taijiro

× Sueda, Taijiro

en Sueda, Taijiro

Search repository
Matsuura, Yuichiro

× Matsuura, Yuichiro

en Matsuura, Yuichiro

Search repository
アクセス権
アクセス権 open access
アクセス権URI http://purl.org/coar/access_right/c_abf2
主題
主題Scheme Other
主題 Artificial heart-lung machine
主題
主題Scheme Other
主題 Low power He-Ne laser
主題
主題Scheme Other
主題 Erythrocyte deformability
主題
主題Scheme Other
主題 Free hemoglobin
主題
主題Scheme NDC
主題 490
内容記述
内容記述 We studied the protective effects of low power He-Ne laser irradiation on erythrocytes against the damage caused by experimental artificial heart-lung machines.   The erythrocyte suspension in a closed circuit was perfused with a roller pump. One of two circuits was used for laser treatment and the other for control. The laser group demonstrated significantly higher erythrocyte deformability and erythrocyte ATP levels, and lower free hemoglobin and lipid peroxide levels. A laser output power of 8.5 mW was most effective in protecting the erythrocytes. A further morphological finding using a scanning electron microscope showed fewer echinocytes and more discocytes in the laser group. This finding was consistent with the results of physiological and biochemical tests.   Low power He-Ne laser irradiation was shown to protect human erythrocytes from the damage caused by an experimental artificial heart-lung machine, raising the possibility of the clinical application of low power He-Ne laser treatment during extracorporeal circulation in cardiovascular surgery.
言語 en
内容記述
内容記述タイプ Other
内容記述 This study was presented at the 6th Japan Laser Therapy Association, held on June 25-26, 1994 in Kobe, at the 32d Japanese Society for Artificial Organs, held on October 12-13, 1994 in Tokyo, at the 47th Japanese Association for Thoracic Surgery, held on October 25-27, 1994 in Chiba, at the 95th Japan Surgical Society, held on April 10-12, 1995 in Nagoya, and at the 33d Japanese Society for Artificial Organs, held on November 7-9, 1995 in Osaka.
出版者
出版者 Hiroshima University Medical Press
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ departmental bulletin paper
出版タイプ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
収録物識別子
収録物識別子タイプ ISSN
収録物識別子 0018-2052
収録物識別子
収録物識別子タイプ NCID
収録物識別子 AA00664312
開始ページ
開始ページ 15
書誌情報 Hiroshima Journal of Medical Sciences
Hiroshima Journal of Medical Sciences

巻 45, 号 1, p. 15-22, 発行日 1996-03
旧ID 37835
戻る
0
views
See details
Views

Versions

Ver.1 2025-02-23 06:31:49.270935
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3